Keywords: Upgrade, SVV, Signals VitroVivo
Question:
What are the exact steps I need to take as the documentation is confusing me?
Answer:
In order to upgrade from SVV3.2 to SVV3.3HF1 by following the steps below:
- Install packages from Signals Inventa 3.3 ONE BY ONE and IN ORDER from \SignalsInventa_3.2_installer\Spotfire Distribution:
- LeadDiscovery-8.6.0.477.spk
- LeadDiscoveryChemCharts.sdn
- PerkinElmer.Signals.Analytics.Apps_4.7.7022.0.spk
- PKISignalsClientNET-3.2.0.118.spk
- SarSolutions-1.1.4.561.spk
- Spotfire.Dxp.PKIUtility_6.0.0.229.spk
- Save
- As part of the upgrading process from VitroVivo 3.2 to VitroVivo 3.3, the administrator should manually remove from the deployment area the following modules:
- PerkinElmer.Signals.InVivo.StudyDesigner
- PerkinElmer.Signals.InVivo.DataModelDesigner
- PerkinElmer.Signals.InVivo.DosePreparation
- PerkinElmer.Signals.InVivo.DoseVerification
- PerkinElmer.Signals.InVivo.MassSpec
- PerkinElmer.Signals.InVivo.SequenceOfEvents
- PerkinElmer.Signals.InVivo.BusinessRule
- PerkinElmer.Signals.InVivo.BaselineCapture
- PerkinElmer.SPR.DataImportApp
- PerkinElmer.SPR.KineticAnalysisApp
- PerkinElmer.SPR.PLAApp
- PerkinElmer.SPR.QAQCApp
- PerkinElmer.SPR.ReferencingApp
- save
- Install (SVV3.3) PerkinElmer-VitroVivo.sdn
- Save
- Install (SVV3.3HF1) PerkinElmer.Signals.Analytics.CalculationsPanel_2.0.0.3846
- save
- Validate
This should complete the upgrade for you from Signals VitroVivo 3.2 HF-03 to Signals VitroVivo 3.3 HF1
Comments
0 comments
Article is closed for comments.