Date Posted:
Product: ChemBioOffice Enterprise
Product: ChemBioOffice Enterprise
Problem:
Why is Barcode Report1 hanging when I try to print a barcode in Inventory Manager?
Solution:
Answer: There may be a bad reference to the ChemDraw 6.0 control in the report. To fix this issue:
- Open cheminv_reports.mdb, found in your ChemOffice Data folder (by default: C:\Chemoffice_data\ChemInv).
- Click Alt-F11 to open the report in the VBA editor.
- Select Tools>References.
- If the CS ChemDraw Control 6.0 Library is checked, select it to uncheck the box.
- Click OK and close cheminv_reports.mdb.
Comments
0 comments
Article is closed for comments.