In Harmony 4.9 it is not possible to establish a connection to a Image Artist system to allow file transfer.
On attempting to establish a connection you will typically see the error:
Cannot establish connection to Columbus.
Please check the Columbus account information.
There was no endpoint listening at http://ip_address:8081/axis2/services/CCInterface2 that could accept the message.
This is often caused by an incorrect address or SOAP action.
See InnerException, if present, for more details.A Harmony software patch is required, this negates the need for the SOAP Web Service that previously featured in Columbus.
Please raise a case with the HCS Technical Support team and request the patch "4.9.2137.273_Patches\Columbus Transfer Fix".
The patch is not required for Harmony versions 5.0.2012.1957 and higher.