When using Data Import app the following error appears:
'The folder 'Signals Data Import App' does not exist on the server.'
Keywords: Data Import, Signals VitroVivo, preference, warning, apps
Details:
To be able to share input file formats across users, Signals VitroVivo requires that a library folder is configured on the server and the library folder path is defined in the corresponding Signals VitroVivo preference.
- Open Spotfire Analyst as administrator and from the Tools menu, go to Library Administration and create a new library folder at the TIBCO Spotfire Server library to store the input data formats.
- Go to Permissions for selected folder and give Browse+Access+Modify permissions to the folder defined in step 1 to the Signals VitroVivo TIBCO Spotfire user groups.
- Go to Tools > Administration Manager > Preferences and select the user groups that you want to add to the library folder. Then populate the Signals Apps > Data Import > Library Folder preference with the folder path created in step 1.
- To configure different folders for different user groups, repeat steps 1 through 3 for each user group
- Restart Spotfire Analyst
Note: When adding the library path do not include “Library/”, as all paths are defined relative to this folder.