Description
Spectrus modules maintain a history of text used in annotation dialogs (for LC/MS, NMR, UV/IR, etc.). There is no in‑application option to clear this list. The history is stored in the Windows registry under technique‑specific keys and can be reset by deleting those keys.
Solution
- Close all ACD/Labs applications.
- Open the Registry Editor (regedit).
- Navigate to the relevant keys for the data type you want to reset:
- LC/MS:
HKEY_CURRENT_USER\Software\Advanced Chemistry Development (ACD)\Spectrus_2025\MASS\Annotation Dialog History - 1D NMR:
HKEY_CURRENT_USER\Software\Advanced Chemistry Development (ACD)\Spectrus_2025\1D NMR\Annotation Dialog History - 2D NMR:
HKEY_CURRENT_USER\Software\Advanced Chemistry Development (ACD)\Spectrus_2025\2D NMR\Annotation Dialog History - Curve:
HKEY_CURRENT_USER\Software\Advanced Chemistry Development (ACD)\Spectrus_2025\CURVE\Annotation Dialog History - UV/IR:
HKEY_CURRENT_USER\Software\Advanced Chemistry Development (ACD)\Spectrus_2025\UVIR\Annotation Dialog History - Chromatography:
HKEY_CURRENT_USER\Software\Advanced Chemistry Development (ACD)\Spectrus_2025\CHROM\Annotation Dialog History
- LC/MS:
- For each technique where you want to clear the history:
- Right‑click Annotation Dialog History and choose Delete.
- Close the Registry Editor.
- Reopen Spectrus. The annotation dialogs will now have empty or default suggestion lists and will start building a new history.
Comments
0 comments
Article is closed for comments.