Description
QbD tools are not accessible from Tools menu in Spectrus Processor
- LC Method Transfer Assistant
- LC Columns Comparison
- LC Columns Selection
After installation, users do not see these QbD tools under the Tools menu.
Solution
There are two solutions to this issue.
- Delete GRASST.XML file
Admin rights are needed.
Make sure all the ACDLabs programs are closed.
Via File Explorer, go to %AppData%\Advanced Chemistry Development\ACDScriptConfig.
In the opened folder locate and delete the grasst.xml file.
Start Spectrus Processor again and check the Tools menu.
- Where do I find the registry entries for the … LC … tools?
HKEY_CURRENT_USER\Software\Advanced Chemistry Development (ACD)\SMMacro2\ANY
2. Use Macro Organizer
Regular MSS users can do it.
Launch Spectrus Processor.
Click the Organizer button:
or from the File menu choose Macro > Organizer…
In the Macro Organizer window that appears, specify ANY Type and click Add...
Navigate to the ACDLabs installation folder, the following location: \\ACD64\EXAMPLES\SPECTRUS\CHROM\SCRIPTS and perform the same procedure to add 3 QbD tools.
For LC Column Comparison:
Select the COL_COMPARE.FSBIN and click Open.
A new script with default name is added:
In the Name field, type a new name for the script: LC Column Comparison.
Perform all 6 steps mentioned above to see a new line in the list:
Click Add.. again, select the QBD_COLSEL.FSBIN file, and click Open.
Perform the same procedure and click OK.
And one more time click Add.. and bring in the GRASST.FSBIN file.
Click OK twice to close Scripts/Macro Location and Macro Organizer windows and check the Tools menu in Spectrus Processor.
The QbD Tools are back to the Tools menu: