Description
When running Estimate Prediction Accuracy in ChromGenius, some molecules may appear in gray, with no estimated retention time (tR) values. This indicates that the software was unable to generate a reliable correlation equation for those structures under the current option/parameter settings.
As a result:
- tR prediction is not calculated (tR Calc remains blank).
- Overall estimation accuracy for the dataset may be suboptimal.
Solution
- Identify affected molecules:
- In the estimation results, note molecules displayed in gray with no predicted tR.
- Test one such molecule:
- Double-click the structure to open it in ChemSketch.
- Attempt to obtain Results for that structure.
- Confirm that the tR Calc column is blank, indicating prediction failure.
- Adjust parameter settings:
- Reopen ChromGenius options for the estimation/prediction model.
- Start with a limited set of parameters to simplify the model.
- Gradually expand parameters:
- Increase the number of allowed parameters stepwise (for example, add additional descriptors one by one).
- After each adjustment, rerun Estimate Prediction Accuracy and check:
- Whether more adequate correlation equations are found.
- Whether previously gray molecules now receive predicted tR values.
- Pay attention to Residual (should be close to 1) and Standard Deviation (close to 0) after each adjustment.
- Review the new results.
- Bring more structures to the database that are similar to the ones with failed calculated tR value.
Comments
0 comments