Description
Accurate chemical shift referencing is important for comparing spectra and reporting assignments. Spectra are usually referenced to TMS (0 ppm) or to residual solvent peaks with known positions.
Solution
- Identify the reference signal:
- TMS at 0.00 ppm, if present.
- Otherwise, a known residual solvent peak (e.g., CHCl3 at 7.26 ppm in CDCl3).
- Use the manual Reference function and click on the reference peak:
- Enter the correct chemical shift value for that peak.
- Apply the reference; all peaks in the spectrum will shift accordingly.
- Verify that other known peaks (e.g., solvent multiplets, internal standards) now appear close to their expected positions.
- Perform assignments and generate reports based on the referenced spectrum.
Comments
0 comments
Article is closed for comments.