Description
Component identification is often based on comparing both retention time and MS/MS spectra to a reference standard. Spectrus can store and compare both for confident identification.
Solution
- Acquire xC/UV/MS data for both:
- The unknown/sample.
- A reference standard under the same chromatographic conditions.
- Import both datasets into Spectrus and identify the corresponding peaks.
- Compare retention times:
- Check that the sample peak and reference peak match within your method’s tolerance (e.g., ±0.1–0.2 min).
- Compare MS/MS spectra:
- Select both spectra by holding the Shift key and selecting both MS windows.
- Overlay or compare the fragmentation patterns for the same precursor m/z by selecting the Show two spectra in a mirrored format icon on the top toolbar.
- Look for matching fragment peaks and relative intensities.
- If both retention times and MS/MS patterns agree:
- Consider the compound positively confirmed (subject to your SOP criteria).
- Store the reference spectrum and retention time as part of your method or library for future comparisons.